<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>47108</ErrorNumber>
<ErrorFamily>Camera</ErrorFamily>
<ErrorMessage>
CAM: Timed out waiting for strobe field from camera
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The camera did not capture a camera strobe field within the expected timeout
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Check that the GMCE Xilinx is version 1.91 or later.
Make sure that the camera power is connected.
Make sure that the camera video cable is connected to both the LED driver board and
the GMCE.
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
Possible camera fault or loose cable.
</Cause>
<LastModified>
08-Apr-2006
Added.
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:51 2008 PST
</Updated>
<Links>
</Links>
</Error>